The forward and backward scattering parts of a new neutron spectrometer, TOSCA-II, replacing the old TFXA and TOSCA-I, have been successfully installed at the ISIS pulsed neutron source. The results show a significant enhancement in the counting rate due to the larger detector area. The improved resolution (to 1.5÷3% of the energy transfer) as compared with the previous instruments (TXFA 2.5 ÷ 3.5%, TOSCA-I 2÷3.5%) has been achieved by increasing the primary flight path from 12 m to 17 m. A chopper has been added in order to avoid neutron frame overlap and to reduce the fast neutron background. Additional diffraction capability will be installed in the near future.
The final configuration of TOSCA neutron spectrometer
Celli M;Cilloco F;Colognesi D;Zoppi M
2001
Abstract
The forward and backward scattering parts of a new neutron spectrometer, TOSCA-II, replacing the old TFXA and TOSCA-I, have been successfully installed at the ISIS pulsed neutron source. The results show a significant enhancement in the counting rate due to the larger detector area. The improved resolution (to 1.5÷3% of the energy transfer) as compared with the previous instruments (TXFA 2.5 ÷ 3.5%, TOSCA-I 2÷3.5%) has been achieved by increasing the primary flight path from 12 m to 17 m. A chopper has been added in order to avoid neutron frame overlap and to reduce the fast neutron background. Additional diffraction capability will be installed in the near future.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.